| NUCON |
|
|
NUCON International, Inc.
7000 Huntley Road Columbus, OH 43229-1035 U.S.A.
Telephone: 614-846-5710 Fax: 614-431-0858 E-mail sales@nucon-int.com |
 |
|
F-1000TM Series
Test Instruments
|
| NUCON Model SN-10 Aerosol Generator |
The NUCON Model SN-10 produces a uniform particle size distribution
aerosol and can be used with a variety of challenge agents. Volumetric output is linear
and particle size distribution is constant over a 10-100 psig input pressure range. The
SN-10 is the perfect upgrade of the "Laskin Nozzle" aerosol generator.
Testing Applications
|
- Precise, continuous output control reproducibly generates the required aerosol particle
size distribution for HEPA filter testing.
- Capable of generating aerosol from any of the currently used liquids (Polyalphaolefin
(PAO), Dioctylphthalate, Dioctylsebacate, Corn Oil, Mineral Oil, Glycol
- The particle size distribution meets the requirements of ASME N510.
- The generator can be sterilized for medical applications.
- Integral strainer to prevent contamination of the nozzle.
- Integral pressure relief valve.
- Integral pressure regulator for output control.
- Available in various capacities.
|
|
Air Pressure vs DOP Output |
|
|
|
|
|
 |
|
Disassembled Generators and Carrying Case
Technical Specifications and Information: |
Size:
15" x 8" (38 cm x 20 cm)
Weight: 9 lb. (4 Kg empty)
Output: Up to 4 g/min of aerosol (More than the output of 12 standard
Laskin Nozzles)
Output Control: Integral pressure regulator
Reservoir: 0.53 gallons (2 liters)
Utility Requirement: Maximum 20 SCFM (550 lpm) Compressed gas (air) at
100 psig (700 kPa) |
Aerosol
size: |
99%
less than 2.0 micro meter
97% less than 1.0 micro meter
90% less than 0.7 micro meter
75% less than 0.5 micro meter |
 |
The "SN" is also available in a smaller size (SN-5). Aerosol
size distribution is identical to the SN-10
Technical Specifications and Information:
- Size: 5" x 10"
- Weight: 4 lb. (empty)
- Output Control: Integral pressure regulator
- Reservoir: 6 Oz. (180 ml.)
- Utility Requirement: 1-10 SCFM at 5-50 psi compressed air
|
 |
NUCON F-1000
Series Test Instruments
|
- Model F Aerosol Generator (Thermal)
- Model F Halide Generator
- Model SN-10 Aerosol Generator (Pneumatic)
|
- Model F Aerosol Detector
- Model SAE Halide Detector
- Model PD-C Pressure Detector
- Model HGP-H Pulse Mode Halide Generator
|
|