F-1000™ Series Test Instruments  
NUCON  

Model F Thermal Aerosol
(DOP) Generator

MAY BE MODIFIED
TO WORK WITH
POLYALPHAOLEFIN

(PAO)

For In-Place Aerosol Leak Testing of
HEPA Filtration Systems

F-1000-DG
Model F Aerosol Generator

Testing Applications

  • Nuclear air and Gas Treatment Systems
  • HEPA Filter Systems
  • Clean Rooms
  • Aerosol Trace Studies

The NUCON Model F is a high-output portable thermal aerosol (DOP) generator. It can provide up to 140 g/min of submicron aerosol.

  • All operation is front panel controlled
  • Designed for field use
  • Fail-safe heater control
  • Leak-tight, long life and safe aerosol generation
SAE-02a.jpg (7805 bytes)

The NUCON F-1000-DG Model F is unique in its ability to produce large volumes of submicron aerosol in a compact, portable instrument. All critical parameters (temperature, pressure and flow) are precisely controlled for consistent, repeatable performance.
Temperature is controlled by solid-state power controllers with factory set curie point temperature sensors. Flow is set using a built-in liquid flowmeter with a needle valve. A separate quarter turn off/on valve is used to start and stop liquid supplied to the heater without changing the flow rate setting.

A dial thermometer displays the heater block temperature. System pressure and reservoir level and all controls are on the front panel.
Liquid to gas phase conversion takes place in a continuous stainless steel coil cast in an aluminum block held in a stainless steel container. Heater cartridges, thermal sensors, and the dial thermometer are inserted in the block. This ensures maximum thermal conduction and mass, along with maximum safety.
Technical Specifications and Information:
Output:
Up to 140 g/min of aerosol equal to concentration of 50 micrograms/liter in an airflow of 85,000 SCFM (144,500 m3/hr) with Main Auxiliary Power ON.

Up to 70 g/min of aerosol equal to concentration of 50 micrograms/liter in an airflow of 42,500 SCFM (42,500 m3/hr) with only Main Power ON.
Aerosol Size:
99 + % less than 3.0 microns

50 + % less than 0.7 micron
10 + % less than 0.4 micron

Reservoir Capacity:
1.4 Liters

Size:
16" (41cm) W

x 9" (23 cm) L
x 15" (38 cm) H

Weight:
35 lbs. (16 Kg)

Power Requirements:
2000 Watts, 120VAC ±10%, 50/60 Hz. Main Power
4000 Watts with Auxiliary Power On

2 scfm Compressed Air
(57 l/min) at
35-115 psig (241-793 kPa)


NUCON F-1000™
Series Test Instruments

  • Model F Aerosol Generator (Thermal)
  • Model F Halide Generator
  • Model SN-10 Aerosol Generator (Pneumatic)
  • Model F Aerosol Detector
  • Model SAE Halide Detector
  • Model PD-C Pressure Detector
  • Model HGP-H Pulse Mode Halide Generator